This simple to operate device consists of two high precision air bearings which both are equipped with high-resolution linear-scales. The measuring values can be displayed in the digital display, and optionally also can be evaluated in the software P-soft.
For the support the nut will be fixed in a high precision prismatic adaption. Nuts with flanges can be adapted on the precise front face with special brackets. Thanks to the stable and homogenous design very low measurement uncertainties within sub-micrometre can be achieved.
Measurement procedure:
Changing the appropriate measurement probe
Setting the height of the measuring head
Fixing the nut in the prism
Moving with the probe into the nuts first circuit and lower the probe
Set zero at the display unit or/and input the value into the software by triggering
Lift the probe with the pneumatic valve and probing the next circuit
TypeDimensions (L×W×H)Inner Ø of the nutMax. length of the nutTravelWeightRequired compressed airAir consumptionElectrical connectionAmbient temperatureMeasurement readingSystem accuracyStraightness horizontal (at probe head)Resolution display unit
The system accuracy applies to a nut lead angle of max. 10°.
The accuracy values apply to 20 °C and stable temperature conditions (ΔT ≤ 0.2K / 24h).
Technical data are subject to change.The data presented are from already implemented test devices. For customized solutions in various dimensions we are your specialist. Please contact us: Contact.